Model No: SX25-AEP8001
Features:
-
8 sampling channels, 16 incubation boards, and 3 washers.
-
Utilizes pressure-type liquid detection.
-
Includes TIP and clot detection functions.
-
High-precision pipetting with a minimum volume of 5μl.
-
Position sensing for real-time monitoring of gripper status.
-
Adjustable multi-spacing and multi-strength grippers, suitable for handling microplates and plate covers of varying sizes.
-
Dual thread control, enabling simultaneous operation of the sampling arm without interference.
Technical Parameters:
Sample & Reagent Unit:
- Sample Racks: 12*20 push-pull sample rack with detection
- Sampling Channel: 8 channels
- Liquid Level Detection: Air pressure level detection, needle blocking intelligent detection
- TIP: 5 TIP racks
- Pipetting: 5~1000μl, capable of parallel sampling
- Pipetting Precision:
- 100μl: CV ≤ 1.0%, accuracy ≤ 2.5%
- 1000μl: CV ≤ 0.5%, accuracy ≤ 1.0%
- Reagent Position:
- 6 reagent positions
- 40 quality control positions
- 12 reagent bottles
- 25 reagent tanks (reagent tank with code identification)
Washing Unit:
- Washer: 3 units with on-board detection function
- Washing Head: 16-channel (32-pin), double-row cleaning, hanging installation, with leakage recovery
- Injection Uniformity: CV < 1.5%
- Washing Residual: < 1μl (residual volume per well)
Reading Unit:
- Reading Channel: 8 independent photometric channels
- Linear Range: 0.000~3.000 Abs
- Repeatability: CV ≤ 1.0%
- Stability: ≤ ±0.003 Abs
- Optical Filters: 4 standard filters (405, 450, 492, 630nm), total of 10 filter positions
- Reading Accuracy:
- ≤ ±0.02 Abs from 0.0 to 1.0 OD
- ≤ ±0.03 Abs from 1.0 to 2.0 OD
- Sensitivity: ≥ 0.01 mg/L
- Channel Difference: ≤ 0.02 Abs
Incubation Unit:
- Incubators: 16 shaking incubators
- Cover Positions: 4 (16 covers)
- Heating Method: Dry heating with cover
- Shaking Function: Horizontal circular oscillation, 1~3 levels adjustable, amplitude 2.5mm
- Temperature Range: RT ~ 60°C
- Temperature Uniformity: ±0.5°C
- Temperature Fluctuation: ±0.5°C
Data Management:
- Software: Compatible with Windows 7 or higher
- LIS System: Bi-directional, supports HL7 protocol
- Results: Absorbance and results reviewable by software
- Printer: External optional
- Built-in Scanner: Optional
Other Functions:
- Biosafety Protection: Optional (Negative pressure working zone, HEPA filter)
- Disinfection: UV lamp
- Lighting: LED
- IAP Function: Firmware can be upgraded online at any time
Working Conditions:
- Power Supply: AC 100~240V, 50/60Hz
- Temperature: 10~40°C
- Humidity: 30~80%
Size & Weight:
- External Size (W * D * H): 1925 * 824 * 1660 mm
- Package Size (W * D * H) mm:
- Machine: 2095 * 950 * 1160 mm
- Base Stand: 2095 * 950 * 1005 mm
- Accessories: 1560 * 930 * 530 mm
- Net Weight (kg):
- Machine: 320
- Base Stand: 153
- Accessories: 68
- Gross Weight (kg):
- Machine: 420
- Base Stand: 245
- Accessories: 122
Reviews